Kit Contents
Kits include a solution of PI in PBS. The PI reagent should be stored protected from light at 2 – 8oC in an airtight container.
Assay Size | Number of Tests |
---|---|
0.25 mL | 50 |
1.5 mL | 300 |
Sample Volume and Chamber Height
The required sample volume for the CellDrop depends on the height of the measurement chamber, which is set in the counting protocol.
Standard Magnification (FLi)
Gap Height (um) | Volume (uL) | Minimum Density (cells/mL) | Maximum Density (cells/mL) |
---|---|---|---|
400 | 40 | 7.0E+02 | 3.1E+06 |
100 | 10 | 2.9E+03 | 1.3E+07 |
50 | 5 | 5.9E+03 | 2.5E+07 |
Gap Height (um) | Volume (uL) | Minimum Density (cells/mL) | Maximum Density (cells/mL) |
---|---|---|---|
400 | 40 | 4.3E+03 | 2.6E+07 |
100 | 10 | 1.7E+04 | 1.0E+08 |
50 | 5 | 3.4E+04 | 2.1E+08 |
Best Practices
- Ensure that the upper and lower chamber surfaces are clean prior to loading sample.
- Lower the arm prior to dispensing sample into the measurement chamber.
- Mix the cell suspension well immediately prior to loading sample, and avoid introducing air bubbles.
- Follow the image guides to adjust focus and fluorescence exposure.
- Adjust the exposure in the red fluorescence channel so that fluorescent cells are not under or overexposed, as shown in the info dialog under the exposure menu.
- Allow cells to settle and stop moving across the live preview before pressing the Count button.

Sample Prep
- Mix cell suspension well. Allow PI to equilibrate to room temperature and vortex briefly.
- Mix PI and cell suspension together in a 50% solution (1 part PI + 1 part cell suspension = Dilution Factor of 2).
- Note: There is no incubation time required. Fluorescence may start to fade if cells are in PI for more than 30 minutes.
- Mix sample thoroughly prior to loading onto the CellDrop.
Sample Measurement
- With the CellDrop arm in the down position, launch the PI app.
- Set sample name, information and protocol as appropriate.
- Pipette well-mixed cells + PI solution and dispense appropriate sample volume into the measurement chamber, using the groove on the lower sample surface as a pipetting guide.
- Note: The volume of sample required depends on the protocol settings for the chamber height. The required volume is displayed on the Count button.
- Adjust focus in the brightfield channel according to the image guide.
- Switch to the red channel and adjust exposure according to the image guide.
- Allow cells to settle, then press the Count button.
Refer to Technical Note 186 – CellDrop Best Practices for additional guidance.
Refer to denovix.com/sds for safety data sheets for CellDrop Cell Counting Assays.